Mechanical Micro pipette, Single & Multi-channel, Adjustable and Fixed volume
MicroPette plus
MicroPette plus multi-channel
Features
• Dispensing head rotates for effortless pipetting convenience
• Individual piston and tip cone assembly allow for easy repair and maintenance
• Compound material-made tip cone secures high sealing performance
• Compatible with most universal tip brands
• Online calibration is available
Fully autoclavable
The MicroPette Plus pipettes are fully autoclavable, enable easy cleaning and reduce the risk of contamination. Steam autoclaving can be performed at 121°C, 1 bar for 20 minutes. After autoclaving, the pipette must be cooled down and left to dry for 12 hours before use.
It is recommended to check the performance of the pipette after each autoclaving.
Greasing and sealing of the pipette piston after every 10th autoclaving will secure an enhanced function.
MicroPette
Features
• Pipettes cover a volume range from 0.1μL to 10mL
• Ergonomic design provides excellent operating experience
• Large display window allows for easy volume identification
• Easy calibration and maintenance
• Each MicroPette supplied with an individual calibration certificate according to ISO8655
• Online calibration is available
MicroPette multi-channel
Features
• 8 and 12 channel pipettes are appropriate for 96 well plates
• Dispensing head rotates for effortless pipetting convenience
• Individual piston and tip cone assemblies allowing easy repair and maintenance
• Compound material-made tip cone se-cures high sealing performance
• Compatible with most universal tip brands
• Online calibration is available
Calibration
All pipettes have been quality tested according to ISO8655-2:2002 and are supplied with individual calibration certificates. The quality control includes gravimetric testing of each pipette with distilled water at 22°C Our website allows users to access online calibration software and achieve accurate and timely calibration.
Online calibration software is FREE to pipette users
Specifications
Volume Range |
Increment |
Test Volume(μl) |
Accuracy error |
Precision error |
||
% |
μl |
% |
μl |
|||
0.1-2.5μl |
0.05μl |
2.5 |
2.50% |
0.0625 |
2.00% |
0.05 |
1.25 |
3.00% |
0.0375 |
3.00% |
0.0375 |
||
0.25 |
12.00% |
0.03 |
6.00% |
0.015 |
||
0.5-10μl |
0.1μl |
10 |
1.00% |
0.1 |
0.80% |
0.08 |
5 |
1.50% |
0.075 |
1.50% |
0.075 |
||
1 |
2.50% |
0.025 |
1.50% |
0.015 |
||
2-20μl |
0.5μl |
20 |
0.90% |
0.18 |
0.40% |
0.08 |
10 |
1.20% |
0.12 |
1.00% |
0.1 |
||
2 |
3.00% |
0.06 |
2.00% |
0.04 |
||
5-50μl |
0.5μl |
50 |
0.60% |
0.3 |
0.30% |
0.15 |
25 |
0.90% |
0.225 |
0.60% |
0.15 |
||
5 |
2.00% |
0.1 |
2.00% |
0.1 |
||
10-100μl |
1μl |
100 |
0.80% |
0.8 |
0.15% |
0.15 |
50 |
1.00% |
0.5 |
0.40% |
0.2 |
||
10 |
3.00% |
0.3 |
1.50% |
0.15 |
||
20-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
20 |
3.00% |
0.6 |
1.00% |
0.2 |
||
50-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
50 |
1.00% |
0.5 |
0.40% |
0.2 |
||
100-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
100 |
2.00% |
2 |
0.70% |
0.7 |
||
200-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
200 |
0.90% |
1.8 |
0.30% |
0.6 |
||
1000-5000μl |
50μl |
5000 |
0.50% |
25 |
0.15% |
7.5 |
2500 |
0.60% |
15 |
0.30% |
7.5 |
||
1000 |
0.70% |
7 |
0.30% |
3 |
||
2-10ml |
0.1ml |
10ml |
0.60% |
60 |
0.20% |
20 |
5ml |
1.20% |
60 |
0.30% |
15 |
||
2ml |
3.00% |
60 |
0.60% |
12 |
8 Chnnel adjustable volume pipette |
||||||
Volume Range |
Increment |
Test Volume(μl) |
Accuracy error |
Precision error |
||
% |
μl |
% |
μl |
|||
0.5-10μl |
0.1μl |
10 |
1.50% |
0.15 |
1.50% |
0.15 |
5 |
2.50% |
0.125 |
2.50% |
0.125 |
||
1 |
4.00% |
0.04 |
4.00% |
0.04 |
||
5-50μl |
0.5μl |
50 |
1.00% |
0.5 |
0.50% |
0.25 |
25 |
1.50% |
0.375 |
1.00% |
0.25 |
||
5 |
3.00% |
0.15 |
2.00% |
0.1 |
||
50-300μl |
5μl |
300 |
0.70% |
2.1 |
0.25% |
0.75 |
150 |
1.00% |
1.5 |
0.50% |
0.75 |
||
50 |
1.50% |
0.75 |
0.80% |
0.4 |
||
12 Chnnel adjustable volume pipette | ||||||
Volume Range |
Increment |
Test Volume(μl) |
Accuracy error |
Precision error |
||
% |
μl |
% |
μl |
|||
0.5-10μl |
0.1μl |
10 |
1.50% |
0.15 |
1.50% |
0.15 |
5 |
2.50% |
0.125 |
2.50% |
0.125 |
||
1 |
4.00% |
0.04 |
4.00% |
0.04 |
||
5-50μl |
0.5μl |
50 |
1.00% |
0.5 |
0.50% |
0.25 |
25 |
1.50% |
0.375 |
1.00% |
0.25 |
||
5 |
3.00% |
0.15 |
2.00% |
0.1 |
||
50-300μl |
5μl |
300 |
0.70% |
2.1 |
0.25% |
0.75 |
150 |
1.00% |
1.5 |
0.50% |
0.75 |
||
50 |
1.50% |
0.75 |
0.80% |
0.4 |
Fixed volume pipettes
Volume Range |
Increment |
Test Volume(μl) |
Accuracy error |
Precision error |
||
% |
μl |
% |
μl |
|||
5μl |
- |
5μl |
1.3% |
0.065 |
1.2% |
0.06 |
10μl |
- |
10μl |
0.8% |
0.08 |
0.8% |
0.08 |
20μl |
- |
20μl |
0.6% |
0.12 |
0.5% |
0.1 |
25μl |
- |
25μl |
0.5% |
0.125 |
0.3% |
0.075 |
50μl |
- |
50μl |
0.5% |
0.25 |
0.3% |
0.15 |
100μl |
- |
100μl |
0.5% |
0.5 |
0.3% |
0.3 |
200μl |
- |
200μl |
0.4% |
0.8 |
0.2% |
0.4 |
250μl |
-- |
250μl |
0.4% |
1.0 |
0.2% |
0.5 |
500μl |
- |
500μl |
0.3% |
1.5 |
0.2% |
1.0 |
1000μl |
- |
1000μl |
0.3% |
3.0 |
0.2% |
2.0 |
2000μl |
- |
2000μl |
0.3% |
6.0 |
0.15% |
3.0 |
5000μl |
- |
5000μl |
0.3% |
15 |
0.15% |
7.5 |