CO2 incubators are widely used in scientific research to grow and maintain cell cultures. A Heal Force CO2 incubator provides you with unsurpassed natural simulation to ensure optimum growth conditions for your culture at all time. That’s why they become the first choice of researchers in fields of application include tissue engineering, in vitro fertilization, neuroscience, cancer research and other mammalian cell research.

Product Detail

Product Tags

2

Safe for cultivation

Cell cultivation in particular is a highly sensitive process in which bacteria, viruses, fungal spores and mycoplasmas can destroy valuable cultures or distort test results, causing more work. Heal Force solves this problem using a unique design and effective method to ensure sterile conditions.
2

90℃ moist heat disinfection (HF90 & HF240)

HF90 and HF240 are equipped with 90℃ moist heat disinfection system. The validated overnight sterilization cycle ensures reliable destruction of germs that could interfere with your work and requires no extra work, such as removal of interior fittings. Mycoplasma is 100% eliminated in a routine disinfection cycle.

Ultraviolet disinfection (HF151UV & HF212UV)

A long-life ultraviolet lamp is equipped at the inner back of HF151UV and HF212UV to sterilize chamber air and water in the reservoir to maintain contamination-free conditions within the chamber. To take maximum effect of disinfection, the wavelength of UV light is kept at 254nm.
2
2

Easy-to-clean design

The cleaning process is significantly simplified by Heal Force's unique, seamless, deep-drawn interior chamber, which reduces any areas where contamination could accumulate. Heal Force incubators offer the best usable-space-to-volume ratio due to the total absence of any additional fittings in the interior chamber.

Inlet filter for CO2 supply

All gas injection lines are filtered via HEPA filter to remove impurities and contaminants before being injected into the chamber. The HEPA filter is able to filter particles larger than 0.3μm at 99.998%.
2
2

Absolutely condensation-free, even at high air humidity level

The high air humidity prevents cell cultures from drying out and also keeps the osmolarity constant in the culture medium. With our CO2 incubators, you can work with air humidity up to 95% while the interal walls remain completely dry ( In order to prevent contamination, however, no condensation must occur). The patented tilted water reservoir system keeps the air humidity absolutely stable.

Optimum temperature control

A reliable air jacketed heating system combined with PT1000 temperature sensors ensures high precision with homogenous heat distribution in the interior. Outstanding dynamics ensure short recovery times and balance out any fluctuations caused by door open for Heal Force CO2 incubators. This provide reliable protection at any time, particularly for sensitive cultures.
2
■ The main heater provides precise temperature control.
■ The bottom heater warms the distilled water and ensures chamber humidity.
■ The outer door heater prevents condensation on the inner door and facilitates quick temperature recovery after door openings.

Divided, inner glass door

Three inner glass doors (HF90) maintains stable climatic conditions, minimizes any changes to the humidity, heat and gas concentration, shortens recovery times significantly and also further reduces the risk of contamination. Six half-size sealed inner glass doors and shelves are optional for model HF240. This makes it possible for several users to work with the same equipment.
2

Auto-start function

The auto-start function, which considerably simplifies the equipment's operation, contains the incubator's automatic start-up and the measuring system's calibration. The thermal conductivity CO2 sensor has its baseline automatically reset without manual adjustment. The incubator can be loaded immediately after the start-up routine is completed.
2

Specifications

Model

HF90

HF240

HF151UV

HF212UV

Construction

 

Exterior dimensions

(W×D×H)

637×762×909(mm)

25.1×30.0×35.8(inch)

780×820×944(mm)

615×768×865mm)

"910×763×795(mm)

30.7×32.3×37.2(inch)

24.2×30.2×34.1(inch)

35.8×30.0×34.1(inch)"

Interior dimensions

(W×D×H)

470×530×607(mm)

18.5×20.8×23.9(inch)

607×583×670(mm)

470×530×607(mm)

"600×588×600(mm)

23.9×22.9×26.4(inch)

18.5×20.9×23.9(inch)

23.6×23.1×23.6(inch)"

Interior Volume

151L/5.3cu.ft.

240L/8.5cu.ft.

151L/5.3cu.ft.

212L/7.5cu.ft.

Net Weight

80kg/176lbs.

80kg/176lbs.

75kg/165lbs.

95kg/209lbs

Interior

Type 304, mirror finish, stainless steel

 

Exterior

Electrolyzed galvanization steel, powder coated

 

Inner door

3 inner doors standard

6 mini inner doors optional

one inner door standard

one inner door standard

Temperature

 

Heating method

Direct Heat & Air Jacket (DHA)

 

Temp. control system

Microprocessor

Temp. sensor

PT1000

Temp. range

5℃ above ambient temperature to 50℃

 

Temp. uniformity

±0.2℃

±0.2℃

±0.2℃

±0.3℃

Temp. stability

±0.1℃

±0.1℃

±0.1℃

±0.1℃

CO2

 

Inlet pressure

0.1 MPa

0.1 MPa

0.1 MPa

0.1 MPa

CO2 control system

Microprocessor

Microprocessor

Microprocessor

Microprocessor

CO2 sensor

Thermal conductivity

Thermal conductivity

Thermal conductivity

Thermal conductivity

CO2 range

0 to 20%

0 to 20%

0 to 20%

0 to 20%

CO2 stability

±0.1%

±0.1%

±0.1%

±0.1%

Humidity

 

Humidifying system

Special designed water reservoir

 

Relative humidity

≥95%

≥95%

≥95%

≥95%

Water reservior volume

3L

3L

4L

6L

Shelves

 

Shelf dimensions

(W×D)

423×445(mm)

16.7×17.5(inch)

423×445(mm)

16.7×17.5(inch)

423×445(mm)

16.7×17.5(inch)

590×510(mm)

23.2×20.1(inch)

Shelf construction

3,10

3,12

3,10

3,12

Standard, Maximum

Type 304, mirror finish, stainless steel

 

Fittings

 

Access port

Standard

Standard

Optional

Optional

Air filter

0.3μm, Efficiency:99.998% (for CO2)

 

Remote alarm contacts

Standard

De-contamination

90℃ moist heat disinfection

90℃ moist heat disinfection

UV lamp

UV lamp

Rated power

600W

735W

600W

700W

Power supply

220V/50Hz (standard)

110V/60Hz (Optional)

Alarm system

Power interruption * High/low temperature * Deviation of CO2 * RH * Door ajar * Independent overheat protection