• lab-217043_1280

Auto 3-part hematology analyzer

Principles

● Impedance method

● Cyanide free colorimetric method


Product Detail

Product Tags

0244

Technical Specification

Parameters

21 Reportable parameters:

WBC, Lym#, Mid#, Gran#, Lym%, Mid%, Gran%, RBC,

HGB,HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV,

PDW, PCT, P-LCR, P-LCC

3 Histograms (WBC, RBC, PLT)

Test Mode

Venous whole blood, Capillary whole blood and Prediluted

Throughput

60 tests/hour

Performance

Parameter
Linearity Range
Carry Over
CV
WBC
0-300x109 /L
≤0.5%
≤2.0%
RBC
0-8x1012 /L
≤0.5%
≤1.5%
HGB
0-250g/L
≤0.5%
≤1.5%
PLT
0-3000x109 /L
≤1.0%
≤4.0%

Sample Volume

≤10ul

Data Memory

Up to 100,000 results (including histograms and patient information)

Display

10.4 inch touch screen

Printout

Built-in thermal printer or External printer

Interface

● 4 USB, 1 LAN, 1 RS-232

● Support LIS connection

● RF card for reagent closing system(optional)

Size/Weight

L * W * H = 411*315*416(mm)

Weight: 21kg

Working Environment

● Power: a.c.100-240V, 50/60Hz

● Temperature: 10-30℃

● Humidity: 20-85%

● Air pressure: 70-106KPa

● Working Latitude: ≤3500m